top of page

Product Description

 

The Model RHB-CAC1 is a chilled mirror hygrometer specifically designed to monitor dew point in compressed air systems. It is a complete integrated system consisting of a microprocessor-based instrument, a chilled mirror dew point sensor, a sample flow meter with valve, and the necessary tubing and fittings. The wall-mounted enclosure allows for easy installation on any vertical surface.

 

Technical Specifications

 

AlarmAudible: >98 dB @ 2' ; Relay: 1 form “C”, non-latching; 10 A @ 240 Vac, 8 A @ 24 Vdc, 1 ⁄2 hp @ 240 Vac ; Visual: Flashing “ALARM” message
Dew/Frost Point Accuracy0.2°C (± 0.5°F)
Dew/Frost Point Temperature Max50 °C
Dew/Frost Point Temperature, Min-40 °C
Display TypeLED
Display Type NoteRed LED, 0.5" tall; 8-digit alphanumeric
Electrical Output4 to 20 mA
EnclosureAluminium
Enclosure RatingsIP55, NEMA Type 12
GrossWeightValue0.15
Humidity SensorFan Cooled 2-stage chilled mirror
Measurement TypeDew Point
MountingSpace-Saving Wall Mount
Operating Temperature, Max50 °C
Operating Temperature, Min0 °C
Output Detail4 to 20 mA (0 to 5 Vdc), -50 to 50°C (-58 to 122°F) or -60 to 75°C (-76 to 167°F)
Process Connection Size1/4 in
Process Connection TypeCompression Fitting
Process Temperature, Max50 °C
Process Temperature, Min-40 °C
Sample Flow Rate0.5 to 5.0 scfh, Integral flow meter
Sample Pressure0 to 150 psig (higher pressures available, consult Engineering Department)
Series IDRHB-CAC1S-Series
Supply Voltage90 to 250 Vac
Temperature Accuracy± 0.2 °C
Temperature Accuracy Note± 0.2 °C (± 0.5 °F)
Temperature Sensor TypeNone
Country of OriginUnited States

 

Contact Us

Catalogue

 

RHB-CAC1S2C - Omega - Air Dewpoint Monitor

SKU: RHB-CAC1S2C
₹1,037,980.00Price
Excluding Sales Tax
0/500
Quantity
9 Weeks
  • Return option is available only for manufacturing defects 

  • One year from the date of Supply

No Reviews YetShare your thoughts. Be the first to leave a review.

Contact us

logo

© 2035 by Trias Automation. Powered and secured by Wix 

bottom of page