When selecting a load cell, one of the most important specifications you will come across is mV/V, such as 1 mV/V, 2 mV/V or 3 mV/V.


But what does mV/V actually mean?


Unlike a conventional pressure transmitter or temperature transmitter that may provide a 4–20 mA or 0–10 V output, a load cell typically produces a very small electrical signal in the millivolt range.


The output is specified in millivolts per volt (mV/V) because the load-cell output depends on the excitation voltage supplied to it.


What does 2 mV/V mean?


Suppose a load cell has a sensitivity of:2 mV/V


This means that at its rated full-scale load, the load cell produces approximately 2 mV of output for every 1 V of excitation.


For example:

Excitation Voltage

Full-Scale Output

5 V

10 mV

10 V

20 mV

12 V

24 mV

The calculation is:

Full-Scale Output = Load Cell Sensitivity × Excitation Voltage


Therefore:

2 mV/V × 10 V = 20 mV

A 1000 kg load cell with 2 mV/V sensitivity and 10 V excitation will produce approximately 20 mV at 1000 kg.


Why does the output depend on excitation voltage?


A typical strain-gauge load cell uses a Wheatstone bridge.

Several strain gauges are bonded to the load-cell's elastic element. When a force is applied, the element deforms slightly. This deformation causes changes in the resistance of the strain gauges.

The gauges are connected in a bridge arrangement.

When there is no load, the bridge is ideally balanced and the differential output is close to zero.

When a load is applied, the resistance of the strain gauges changes. This unbalances the bridge and produces a small differential voltage.

The excitation voltage provides the electrical energy for this bridge.

Therefore, increasing the excitation voltage also increases the bridge's output voltage.

This is why load-cell sensitivity is expressed as mV/V rather than simply mV.


Is the load-cell output linear?


Ideally, the output is proportional to the applied load.

For a 2 mV/V load cell with 10 V excitation, the full-scale output is approximately 20 mV.

Therefore:

Applied Load

Approx. Output

0%

0 mV

25%

5 mV

50%

10 mV

75%

15 mV

100%

20 mV

This represents the ideal relationship.


In a real load cell, the output will have small deviations from perfect linearity. Manufacturers therefore specify parameters such as non-linearity, hysteresis, repeatability and zero balance.


Why is mV/V used instead of a fixed mV output?


Consider two identical 2 mV/V load cells.

If one is powered with 5 V:

2 mV/V × 5 V = 10 mV

If the other is powered with 10 V:

2 mV/V × 10 V = 20 mV

The load-cell sensitivity remains 2 mV/V, even though the actual output voltage is different.

This makes mV/V a useful way of specifying the performance of the load cell independently of a particular excitation voltage.


What happens if the excitation voltage changes?


This is an important consideration when designing a weighing or force-measurement system.

Suppose a 2 mV/V load cell is operating with 10 V excitation.

At full load:

2 × 10 = 20 mV

If the excitation falls to 9 V:

2 × 9 = 18 mV

The same mechanical load now produces a different electrical output.

Therefore, a stable excitation supply is important for accurate measurement.

Many load-cell instruments and transmitters use a ratiometric measurement approach, where the excitation voltage is also used as a reference. This helps reduce errors caused by changes in excitation voltage.


Why is the load-cell signal so small?


A load cell is designed to measure very small physical deformation.

As a result, its electrical output is also very small—often only a few millivolts at full scale.


For example:

2 mV/V × 10 V = 20 mV

This is much smaller than the signals normally used by industrial control systems.

Therefore, a load cell is commonly connected to a load-cell amplifier, indicator or transmitter.

The signal can then be converted into an industrial output such as:

mV/V → 0–10 V

or

mV/V → 4–20 mA

or a digital communication signal.


Example: Selecting an amplifier


Consider a 2000 kg load cell with:

  • Capacity: 2000 kg

  • Sensitivity: 2 mV/V

  • Excitation: 10 V

The full-scale bridge output is:

2 mV/V × 10 V = 20 mV

The amplifier therefore needs to accept a load-cell input around 0–20 mV (with appropriate allowance for zero offset and the manufacturer's specified input range).

If the application requires a 4–20 mA output, the transmitter can scale:

0 kg → 4 mA

1000 kg → 12 mA

2000 kg → 20 mA

This allows the very small load-cell signal to be integrated with conventional industrial PLCs, controllers and data-acquisition systems.


Key points to remember


mV/V is the sensitivity of a load cell.

A 2 mV/V load cell produces approximately:

2 mV output for every 1 V of excitation at full load.

Therefore:

Output = mV/V × Excitation Voltage × Load %


For example:

2 mV/V × 10 V × 50% = 10 mV


The output is approximately linear with load, but real-world load cells have small errors that are specified in their datasheets.


Understanding mV/V is therefore essential when selecting a load cell, excitation supply, load-cell amplifier or 4–20 mA transmitter.